The TCBHP-1A tests the magnetic properties of motor stator/rotor cores and wound transformer cores, measuring flux density B, magnetizing force H and iron loss P. Because stamping, stacking and heat treatment considerably change the material properties (especially iron loss) from raw sheet to finished core, verification of the finished core is essential. Improvements over comparable products overcome the large measurement error seen with narrow-yoke stators, while a 16-bit high-speed MCU with 16-bit A/D increases sampling depth and digital filtering efficiency.
Specifications are from the product manual; models and configurations are subject to the latest technical documentation
| Test Object | Motor stator/rotor cores or wound transformer cores |
|---|---|
| Size Range | ID 20 mm – OD 330 mm; yoke width 2–200 mm |
| Stack Height / Magnetic Path | Stack 10–400 mm; magnetic path 10–1000 mm |
| Flux Density B | 0~2.0 T |
| Magnetic Field Strength H | 0~1000 A/m |
| Iron Loss P | 0~20 W/kg |
| Magnetizing Method | AC sinusoidal current magnetization |
| Measurement Accuracy | ±(2.5% + 2 digits) |
| Power Supply | AC 220 V ± 22 V, 50 Hz ± 2 Hz |
| Max. Power Consumption | 2 KW |
| Dimensions · Weight | W 520 × D 400 × H 600 mm, approx. 45 kg |
※ Specifications, appearance and configuration are subject to the latest technical documentation.