The TCIL-2A measures the iron loss of single silicon steel sheets for producers, distributors and manufacturers of motors and transformers. Building on the standard model, it tests irregular samples of varying width and thickness down to a minimum width of only 3.0 mm — ideal for incoming inspection and process monitoring of stamped laminations, offcuts and other non-standard silicon steel parts.
Specifications are from the product manual; models and configurations are subject to the latest technical documentation
| Test Object | Single silicon steel sheet (cold-rolled, hot-rolled, oriented, non-oriented and irregular stamped parts) |
|---|---|
| Size Range | Thickness 0.10–0.99 mm; width ≥3.0 mm; length ≥25 mm |
| Flux Density B | 1.0–1.7 T (continuously adjustable) |
| Iron Loss P | 0~20 W/kg |
| Test Frequency F | 50 Hz / 60 Hz (freely switchable) |
| Magnetizing Method | AC sinusoidal current magnetization |
| Measurement Accuracy | ±5% (correlated to the Epstein frame method) |
| Power Supply | 90 V – 240 V |
| Max. Power Consumption | 35 W |
| Dimensions · Weight | W 210 × D 130 × H 280 mm, approx. 3.5 kg |
※ Specifications, appearance and configuration are subject to the latest technical documentation.
The TCIL-1A measures the iron loss of single silicon ste
View DetailsThe TCIL-5A measures the iron loss of single silicon ste
View DetailsThe TCIL-3A measures the iron loss and flux density of s
View DetailsCopyright © 2026 Changzhou Sicharm Electronic Technology Co., Ltd. All rights reserved. · Su ICP No. 2026088983